首页> 外文期刊>The Journal of arthroplasty >Initial tension loss in cerclage cables
【24h】

Initial tension loss in cerclage cables

机译:环扎电缆的初始张力损失

获取原文
获取原文并翻译 | 示例
           

摘要

Cerclage cables, frequently used in the management of fractures and osteotomies, are associated with a high failure rate and significant loosening during surgery. This study compared the capacity to maintain tension of different types of orthopaedic cable systems. Multifilament Cobalt-Chrome (CoCr) cables with four different crimp/clamp devices (DePuy, Stryker, Zimmer and Smith&Nephew) and one non-metallic Nylon (Ny) cable from Kinamed were instrumented with a load cell to measure tension during insertion. Significant tension loss was observed with crimping for all cables (P<. 0.05). Removing the tensioner led to an additional unexpected tension loss (CoCr-DePuy: 18%, CoCr-Stryker: 29%, CoCr-Smith&Nephew: 33%, Ny: 46%, and CoCr-Zimmer: 52%). The simple CoCr (DePuy) cable system outperformed the more sophisticated locking devices due to its significantly better ability to prevent tension loss.
机译:经常在骨折和截骨术中使用的环扎电缆具有较高的失效率,并且在手术过程中明显松动。这项研究比较了保持不同类型的骨科电缆系统张力的能力。将带有四种不同的压接/夹紧装置(DePuy,Stryker,Zimmer和Smith&Nephew)的多丝钴铬(CoCr)电缆和Kinamed的一根非金属尼龙(Ny)电缆用称重传感器进行测量,以测量插入期间的张力。对于所有电缆,在压接时均观察到明显的张力损失(P <0.05)。卸下张紧器会导致额外的意外张力损失(CoCr-DePuy:18%,CoCr-Stryker:29%,CoCr-Smith&Nephew:33%,Ny:46%和CoCr-Zimmer:52%)。简单的CoCr(DePuy)电缆系统具有明显更好的防止张力损失的能力,其性能优于更复杂的锁定设备。