Assessment of a Medial Pivot Total Knee Arthroplasty Design in a Cadaveric Knee Extension Test Model

机译:在尸体膝盖延伸试验模型中评估内侧枢轴全膝关节置换设计

获取原文
获取原文并翻译 | 示例
           

摘要

A total knee has been designed to mimic less-compliant medial and more-compliant lateral behavior. In vivo testing compared open-kinematic chain behaviors of cadaver knees in their normal state and after implantation of the knee prosthesis. Specimen's limbs were computed tomography scanned, and infrared arrays on tibia and femur were registered to bone markers. Motion of the joint and quadriceps force were reported from 90° flexion to full extension. Less medial and more lateral anterior-posterior motion was seen in both the intact and the implanted knees. Tibiofemoral rotation and translation were similar in direction but were reduced in magnitude for the prosthetic knees. Quadriceps force, defined as that applied force required to extend the knee, required after implantation was variable between specimens but not statistically different from the intact condition. The prosthesis tested exhibits kinematic behavior similar to that in their normal state, with no difference in quadriceps force required for extension.
机译:全膝关节已被设计为模仿较不顺从的内侧和较顺从的外侧行为。体内测试比较了尸体膝盖在正常状态和膝关节假体植入后的开放运动链行为。用计算机断层扫描对标本的四肢进行扫描,并将胫骨和股骨上的红外阵列记录到骨标志物上。据报道,关节和股四头肌的运动从90度屈曲到完全伸展。在完整的膝盖和植入的膝盖中,内侧和外侧的前后运动较少。胫股旋转和平移的方向相似,但假膝的幅度减小。股四头肌力定义为植入后所需的伸膝所需的力,在标本之间是可变的,但与完好无损在统计学上没有差异。所测试的假体表现出与正常状态下相似的运动学特性,四头肌伸展所需的力没有差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号