首页> 外文期刊>The Journal of arthroplasty >Intramedullary canal pressure distribution: an experimental parametric study.
【24h】

Intramedullary canal pressure distribution: an experimental parametric study.

机译:髓腔内压力分布:一项实验性参数研究。

获取原文
获取原文并翻译 | 示例
       

摘要

Intramedullary instrument/device insertion during total joint replacement can generate high intramedullary pressures, with potential for clinical fat embolism syndrome. We evaluated effects of process parameters on intramedullary pressure generated when hammering an implant into a simulated femur analogue. Lower implant insertion speed, lower hammering force, a rubber- compared with steel-tipped hammer and a larger synthetic bone-to-implant radial gap reduced intramedullary pressure generated. Number of hammer strikes required to insert an implant increased with increasing marrow viscosity, whereas intramedullary pressure decreased with increasing viscosity. Significant intramedullary pressure reduction is obtained by applying more strikes of lower peak force using a rubber hammer tip, as opposed to fewer larger strikes. Caution should be exercised when instrumenting the distal femur, where viscosity is low and very high intramedullary pressures can be generated.
机译:全关节置换过程中插入髓内器械/器械会产生较高的髓内压力,可能会导致临床脂肪栓塞综合征。当将植入物锤击成模拟股骨类似物时,我们评估了工艺参数对产生的髓内压的影响。与钢制锤相比,较低的植入物插入速度,较低的锤击力,橡胶和较大的合成骨-植入物径向间隙可减少产生的髓内压力。插入植入物所需的锤击次数随骨髓粘度的增加而增加,而髓内压力随粘度的增加而降低。与较大的冲击相比,通过使用橡胶锤头施加更多的较低峰值力的冲击,可以获得明显的髓内压力降低。当对股骨远端进行手术时应谨慎,因为股骨远端的粘度较低,可能会产生很高的髓内压。