首页> 外文期刊>The Journal of arthroplasty >Clinical performance of contemporary tibial polyethylene components.
【24h】

Clinical performance of contemporary tibial polyethylene components.

机译:当代胫骨聚乙烯组件的临床表现。

获取原文
获取原文并翻译 | 示例
       

摘要

A postclinical retrieval analysis was performed on 43 polyethylene tibial components of a contemporary total knee arthroplasty system with implantation duration between 12 and 80 months. Components were scored for 8 potential modes of surface wear or damage on the top and back surfaces. Moderate backside wear of 4.1 microm/y was documented by measuring the extent of manufacturer's engraved lettering removal. Neither the topside nor backside score correlated with duration of in vivo function. No component experienced topside or backside delamination, cracking, or significant deformation. The greatest contribution to wear and damage score was pitting and scratching secondary to bone cement debris. The extent of both wear and damage experienced by these components was moderate, in comparison with that previously reported with older implant systems.
机译:对当代全膝关节置换系统的43个聚乙烯胫骨组件进行了临床后检索分析,植入时间在12到80个月之间。对组件的表面和背面的8种潜在表面磨损或损坏方式进行评分。通过测量制造商雕刻文字的去除程度,可以发现背面的中等磨损为4.1微米/年。顶部和背面评分均与体内功能的持续时间无关。没有组件出现顶部或背面的分层,开裂或明显变形。磨损和损坏评分的最大贡献是继骨水泥碎片后的点蚀和刮擦。与以前的旧式植入系统相比,这些组件遭受的磨损和损坏的程度是中等的。