首页> 外文期刊>The Journal of arthroplasty >Failure of a metal-on-metal total hip arthroplasty from progressive osteolysis.
【24h】

Failure of a metal-on-metal total hip arthroplasty from progressive osteolysis.

机译:进行性骨溶解导致金属对金属的全髋关节置换术失败。

获取原文
获取原文并翻译 | 示例
           

摘要

Ultra-high-molecular weight polyethylene (UHMWPE) wear, debris-induced osteolysis is a frequent cause of failure of total hip arthroplasty. Metal-on-metal total hip arthroplasty eliminates the generation of UHMWPE particulate debris. Although the volumetric wear of a metal-on-metal articulation may be lower than a metal-UHMWPE articulation, the number of particles may be higher. Osteolysis can develop in response to metallic and UHMWPE debris. The following case of massive osteolysis associated with large amounts of cobalt-chrome wear debris shows adhesive and abrasive wear mechanisms, as well as wear caused by third-body cobalt-chrome debris and impingement of the femoral component against the rim of the acetabular cup, which led to failure of a metal-on-metal total hip arthroplasty.
机译:超高分子量聚乙烯(UHMWPE)磨损,碎片诱导的骨溶解是全髋关节置换术失败的常见原因。金属对金属的全髋关节置换术消除了UHMWPE颗粒碎片的产生。尽管金属对金属铰接的体积磨损可能低于金属-UHMWPE铰接的体积磨损,但颗粒数量可能更高。对金属和UHMWPE碎片的反应会导致骨溶解。以下发生大量钴铬合金磨损碎片而引起的大规模骨溶解事件显示出粘着剂和磨料的磨损机理,以及由第三体钴铬合金碎片和股骨组件撞击髋臼杯边缘引起的磨损,导致金属对金属的全髋关节置换术失败。