A unique, straight-stemmed, proximally porous-coated, modular hip arthroplasty system, coated with thin-film (5- to 9-mum), titanium-nitride ceramic, was used clinically in 130 hip arthroplasties in 117 patients who were followed over a 2- to 12-year interval (mean, 6.45 years). Harris Hip Scores demonstrated 82.3% excellent, 15.4% good, 2.3% fair, and 0% poor results. Thigh pain that limited activities of daily living was seen in 0.8% (1 of 130) hips. Kaplan-Meier survival estimates using an endpoint of revision of any component for any reason demonstrated an overall survival of 95.5% during the 12-year interval. Cementless fixation survivorship of the acetabular and femoral components was 98.5% during the 12-year interval.
机译:一种独特的,直柄,近端多孔涂层的模块化髋关节置换系统,涂有薄膜(5至9毫米)的氮化钛陶瓷,临床用于117例患者的130例髋关节置换术中2到12年的间隔(平均6.45年)。 Harris Hip分数显示了82.3%的优异,15.4%的良好,2.3%的公平,0%的不良结果。在0.8%(130个中的1个)臀部中发现了限制日常生活活动的大腿疼痛。 Kaplan-Meier生存估计值由于任何原因使用任何成分的修订终点,表明在12年的时间间隔内总生存率为95.5%。在12年的时间间隔内,髋臼和股骨组件的非骨水泥固定存活率为98.5%。