Fluoride-Containing Acrylic Bone Cement in Total Hip Arthroplasty. Randomized Evaluation of 97 Stems Using Radiostereometry and Dual-Energy X-ray Absorptiometry.

机译:全髋关节置换术中的含氟丙烯酸骨水泥。使用放射立体法和双能X线吸收法对97个茎进行随机评估。

获取原文
获取原文并翻译 | 示例

摘要

Ninety patients (97 hips) scheduled for total hip arthroplasty were stratified to fixation of the femoral component using fluoride-containing cement or Palacos with gentamicin. Whole polyethylene Reflection and press-fit Trilogy cups were used. All patients received Spectron EF stem. The micromotions of the stem were measured with radiostereometric analysis and the periprosthetic bone mineral density with automatic and manual dual-energy x-ray absorptiometry (DEXA) analysis. At 2 years, the choice of cement did not influence the subsidence or rotations of the stem. The DEXA analysis revealed more loss of periprosthetic bone mineral density in fluoride cement group. We speculate that forming of fluorapatite crystals, toxic effects of the fluoride, or lower radiopacity of the fluoride cement might explain this finding. According to our study with 2-year of follow-up, there is no obvious advantage of addition of fluoride to acrylic bone cement when used to fixate the femoral component in total hip arthroplasty.
机译:计划进行全髋关节置换术的90例患者(97髋)使用含氟水泥或庆大霉素的Palacos固定股骨组件。使用全聚乙烯反射杯和压合Trilogy杯。所有患者均接受Spectron EF茎。通过放射立体分析法测量茎的微动,并通过自动和手动双能X射线吸收法(DEXA)分析来测量假体周围的骨矿物质密度。在2年时,水泥的选择不影响杆的沉陷或旋转。 DEXA分析显示,氟水泥组的假体周围骨矿物质密度损失更多。我们推测氟磷灰石晶体的形成,氟化物的毒性作用或氟化物水泥的较低射线不透性可能解释了这一发现。根据我们为期2年的随访研究,当用于固定全髋关节置换术中的股骨组件时,在丙烯酸骨水泥中添加氟化物没有明显的优势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号