In this article, the hydrogen masers in the NTSC (National Time Service Center) timekeeping laboratory are tested. In order to avoid the impact of larger noise of caesium atomic clocks, TA(k) or UTC(k) is not used as reference, instead, the four hydrogen masers are mutually referred and tested. The frequency stability of hydrogen masers is analyzed using the four-cornered hat method, and the Allan standard deviations of each single hydrogen maser in different sample times are estimated. Then, according to the characteristics of hydrogen masers, by removing the trend term, excluding outliers, and smoothing the data with a mathematical method to separate the Gaussian noises of hydrogen masers, and finally by through the Kolmogorov-Smirnov test, the Gaussian noise of each hydrogen maser is estimated.
机译:在本文中,对NTSC(国家时间服务中心)计时实验室中的氢气脉搏进行了测试。为了避免铯原子钟的较大噪声的影响,不使用TA(k)或UTC(k)作为参考,而是相互参照并测试了四个氢脉石。使用四角帽法分析了氢激射器的频率稳定性,并估计了不同采样时间内每个单个氢激射器的艾伦标准偏差。然后,根据氢气激射器的特征,通过去除趋势项,排除异常值,并使用数学方法对数据进行平滑处理以分离氢气激射器的高斯噪声,最后通过Kolmogorov-Smirnov检验,得出氢激子的高斯噪声。估计每个氢激射器。