首页> 外文期刊>The international journal of medical robotics + computer assisted surgery: MRCAS >Tailor-made surgical guide based on rapid prototyping technique for cup insertion in total hip arthroplasty.
【24h】

Tailor-made surgical guide based on rapid prototyping technique for cup insertion in total hip arthroplasty.

机译:基于快速原型技术的量身定制的手术指南,可在全髋关节置换术中插入杯。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: A surgical guide made by the rapid prototyping (RP) technique for cup insertion in total hip arthroplasty might be useful to avoid malalignment of the cup, which indicates postoperative complications. METHODS: To address this research question, we applied a RP-based guide to 24 patients with their CT images. We designed it to fit onto the acetabular edge and to insert a Kirschner wire (K-wire) which indicated a planned cup direction. We intraoperatively placed it on the acetabular edge, inserted the K-wire through the guide on the superior acetabulum and implanted the cup while observing the alignment of the K-wire. We also recorded the additional time needed to use the guide. RESULTS: The mean cup accuracy between planned and postoperative alignments was 2.8 degrees (SD = 2.1 degrees ) for abduction and 3.7 degrees (SD = 2.7 degrees ) for anteversion. The mean additional time was 3.5 (range 2-6) min. CONCLUSION: We can use this guide with acceptable accuracy and without consuming an excessive amount of time.
机译:背景:通过快速原型(RP)技术制作的用于在全髋关节置换术中插入杯的手术指南可能有助于避免杯错位,这表明术后并发症。方法:为解决该研究问题,我们对24例患者的CT图像应用了基于RP的指南。我们将其设计为适合髋臼边缘并插入表示计划的杯方向的克氏针(K线)。我们在手术中将其放置在髋臼边缘,将K线穿过上髋臼上的导向器插入,并在观察K线对齐的同时植入杯。我们还记录了使用指南所需的额外时间。结果:计划的和术后的对齐之间的平均杯精度外展为2.8度(SD = 2.1度),前倾为3.7度(SD = 2.7度)。平均额外时间为3.5分钟(范围2-6)。结论:我们可以以可接受的准确性使用本指南,而不会花费过多时间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号