Validation of a single-channel airflow monitor for screening of sleep-disordered breathing.

机译:验证用于检查睡眠呼吸障碍的单通道气流监测器。

获取原文
获取原文并翻译 | 示例
           

摘要

A simple screening method for sleep-disordered breathing (SDB) is desirable for primary care practices. In the present study, a simple monitor, which utilises a new type of flow sensor and a novel algorithm, was prospectively validated. Home recording for 2 nights with the monitor only, followed by in-laboratory recording with the monitor together with polysomnography, were carried out in consecutive patients (n = 100) suspected of SDB. A subjective sleep log was also recorded. The signal was analysed using power spectral analysis, which yielded the flow respiratory disturbance index (flow-RDI). There was no recording failure at home. The reproducibility of the flow-RDI between the 2 nights at home was high (intraclass correlation coefficient = 0.92). The sensitivity and specificity of the in-laboratory flow-RDI to diagnose SDB were 0.96 and 0.82, 0.91 and 0.82, and 0.89 and 0.96, for apnoea/hypopnoea index (AHI) > or =5, > or =15 and > or =30 events x h(-1), respectively. The diagnostic ability in low-severity subgroups (female, normal weight, AHI <15 events x h(-1)) was almost comparable to that in the entire group. Excluding subjective waking time on the sleep log from the recording time had no significant effect on the flow-RDI. The single-channel monitor is considered feasible for ambulatory sleep disordered breathing monitoring because of its easy applicability, high reproducibility and relatively high agreement with polysomnography results.
机译:对于初级保健实践,需要一种用于睡眠呼吸障碍(SDB)的简单筛查方法。在本研究中,前瞻性地验证了使用新型流量传感器和新颖算法的简单监控器。在连续怀疑有SDB的患者(n = 100)中,仅在监护仪的情况下在家进行2晚的记录,然后在监护仪中进行实验室记录和多导睡眠监测。还记录了主观睡眠日志。使用功率谱分析对信号进行分析,得出流动呼吸干扰指数(flow-RDI)。家里没有录制失败。在家两晚之间流RDI的可重复性很高(类内相关系数= 0.92)。对于呼吸暂停/呼吸不足指数(AHI)>或= 5,>或= 15和>或=,实验室内流动RDI诊断SDB的敏感性和特异性分别为0.96和0.82、0.91和0.82、0.89和0.96。 30个事件xh(-1)。低严重度亚组(女性,正常体重,AHI <15事件x h(-1))的诊断能力几乎与整个组相当。从记录时间中排除睡眠日志中的主观醒来时间对流RDI没有显着影响。单通道监测器因其易用性,高可重复性以及与多导睡眠图检查结果相对较高的一致性,被认为可用于动态睡眠呼吸障碍监测。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号