首页> 外文期刊>The Analyst: The Analytical Journal of the Royal Society of Chemistry: A Monthly International Publication Dealing with All Branches of Analytical Chemistry >A novel calibration procedure for trace analytical measurements: application to the analysis of polybrominated diphenyl ethers by GC-MS
【24h】

A novel calibration procedure for trace analytical measurements: application to the analysis of polybrominated diphenyl ethers by GC-MS

机译:痕量分析测量的新型校准程序:通过GC-MS在多溴联苯醚的分析中的应用

获取原文
获取原文并翻译 | 示例
           

摘要

The measurement of polybrominated diphenyl ethers ( PBDEs) by gas chromatography-mass spectrometry ( GC-MS) is highly challenging due to existence of the analytes at trace levels and the instability of the higher molecular weight PBDEs. These properties make accurate analysis especially problematic when utilising a low-resolution 'bench-top' GC-electron impact ( EI) mass spectrometer. We discuss the problems associated with the analyses of PBDEs on these common GC-EI-MS systems and present a novel approach to calibration in these circumstances. We propose that this 'mixed calibration procedure' allows analyses to be performed that would not be possible using traditional calibration techniques, and that the procedure may be applied to a wide range of applications throughout analytical science.
机译:由于气相色谱-质谱联用技术(GC-MS)的多溴联苯醚(PBDEs)的测量具有很高的挑战性,因为存在痕量的分析物以及较高分子量的PBDEs的不稳定性。当使用低分辨率的“台式” GC电子碰撞(EI)质谱仪时,这些属性使准确的分析尤其成问题。我们讨论了与这些常见GC-EI-MS系统上的PBDEs分析相关的问题,并提出了在这种情况下进行标定的新颖方法。我们建议,这种“混合校准程序”允许执行使用传统校准技术不可能进行的分析,并且该程序可以应用于整个分析科学的广泛应用。