首页> 外文期刊>The Analyst: The Analytical Journal of the Royal Society of Chemistry: A Monthly International Publication Dealing with All Branches of Analytical Chemistry >The response of a membrane inlet ion mobility spectrometer to chlorine and the effect of water contamination of the drying media on ion mobility spectrometric responses to chlorine
【24h】

The response of a membrane inlet ion mobility spectrometer to chlorine and the effect of water contamination of the drying media on ion mobility spectrometric responses to chlorine

机译:膜入口离子迁移谱仪对氯的响应以及干燥介质的水污染对离子迁移谱对氯的响应的影响

获取原文
获取原文并翻译 | 示例
           

摘要

The negative mode response fo a membrane inlet ion mobility spectrometer, interfaced to an Extrel-C150 mass spectrometer, to chlorine in dry air over the range of concentrations 0-200 mg m~(-3) was characterised. Three distinct product ions were observed: an assembly of [Cl_2(H_2O)_n]~- and [Cl(H_2O)_n]~- entities with a reduced mobility of 2.19 cm~2 V~(-1) s~(-1), a carbon dioxide based cluster, [Cl_2(CO_2)(H_2O)_3]~- with a reduced mobility of 1.93 cm~2 V~(-1) s~(-1) and a dimer product thought to be [Cl_2]_2(H_2O)]~- with a reduced mobility of 1.70 cm~2 V~(-1) s~(_1). The response to chlorine was found to be sensitive with a limit of detection significantly less than 1 mg m~(-3). Further exposures to chlorine were also undertaken with 5 and 10% by mass of water added to the adsorbent traps used to purify the drift and source gases within the instrument. This was done to simulate ageing of the air purification media. No changes in the chemistry of the product ions were observed with increasing levels of water contamination of the adsorbent media. The sensitivity was reduced and the observed drift times increased with increasing levels of water contamination with the reduced mobilities for the three product ion groups above found to vary across the ranges: 2.10-2.19, 1.84-1.93 and 1.66-1.0 cm~2 V~(-1) s~(-1), respectively.
机译:表征了与Extrel-C150质谱仪对接的膜入口离子迁移谱仪对干燥空气中浓度在0-200 mg m〜(-3)范围内的氯的负模响应。观察到三个不同的产物离子:[Cl_2(H_2O)_n]〜-和[Cl(H_2O)_n]〜-实体的组装,其迁移率降低了2.19 cm〜2 V〜(-1)s〜(-1 ),二氧化碳基团簇[Cl_2(CO_2)(H_2O)_3]〜-,其迁移率降低了1.93 cm〜2 V〜(-1)s〜(-1),二聚体产物被认为是[Cl_2 ] _2(H_2O)]-的迁移率降低了1.70 cm〜2 V〜(-1)s〜(_1)。发现对氯的反应敏感,检出限明显小于1 mg m〜(-3)。还向氯气中添加了质量分数为5%和10%的水,这些水被添加到用于净化仪器中漂移气体和源气体的吸附阱中。这样做是为了模拟空气净化介质的老化。随着吸附介质中水污染水平的提高,未观察到产物离子的化学变化。随着水污染程度的增加,灵敏度降低,观察到的漂移时间增加,并且上述三个产物离子基团的迁移率降低,发现其变化范围为2.10-2.19、1.84-1.93和1.66-1.0 cm〜2 V〜 (-1)s〜(-1)。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号