A biomechanical study of simulated femoral neck fracture fixation by cannulated screws: effects of placement angle and number of screws.

机译:空心螺钉固定模拟股骨颈骨折的生物力学研究:放置角度和螺钉数量的影响。

获取原文
获取原文并翻译 | 示例
       

摘要

The angle of placement of hip screws to fix femoral neck fractures is still a controversial subject, and it must be addressed. In the study reported here, we compared the relative stiffness of fixation of simulated Pauwels type III femoral neck fractures fixed with either 2 or 3 cannulated screws implanted at 135 degrees, 145 degrees, and 150 degrees. Each femur was fixed with 2 or 3 cannulated screws and tested under axial loading and anteroposterior (AP) bending. Then each femur was fatigued to 1000 cycles and tested to failure. Fourteen femurs were tested. Results showed that axial stiffness values were not statistically different at different angles. AP bending stiffness of the high-angle (150 degrees) construct was significantly higher than that of either of the other 2 constructs (for 2 screws only). Two-screw fixation appears to be adequate; adding a third screw may not be necessary.
机译:髋螺钉固定股骨颈骨折的角度仍然是一个有争议的话题,必须解决。在这里报道的研究中,我们比较了用135度,145度和150度植入的2个或3个空心螺钉固定的模拟Pauwels III型股骨颈骨折的相对固定刚度。每个股骨用2或3个空心螺钉固定,并在轴向载荷和前后弯曲(AP)下进行测试。然后将每个股骨疲劳至1000个周期并测试其是否衰竭。测试了十四个股骨。结果表明,轴向刚度值在不同角度上没有统计学差异。高角度(150度)构造的AP弯曲刚度显着高于其他2个构造中的任何一个(仅适用于2个螺钉)。两螺钉固定看来已足够。无需添加第三颗螺钉。