Ceramic-on-ceramic failure secondary to head-neck taper mismatch.

机译:继发于头颈部锥度不匹配的陶瓷对陶瓷的破坏。

获取原文
获取原文并翻译 | 示例
           

摘要

We report a case of catastrophic failure of ceramic-on-ceramic total hip arthroplasty without precipitating event or trauma. The patient was a 64-year-old woman who had degenerative osteoarthritis and underwent the index primary total hip arthroplasty 3 years earlier. Intraoperative findings included an intact ceramic femoral head, a slightly damaged ceramic liner insert, diffuse metallosis, and excessive wear of the trunnion of the stem. After removal of the metallic debris, excision of metalloid tissue, and copious lavage of the joint, the prosthesis was revised to a modular revision system. Although previous operative reports had been reviewed before surgery, there was no indication of a head-neck taper mismatch. Only after revision surgery was performed, and high suspicion arose, were previous implant records analyzed and the mismatch identified.
机译:我们报道了陶瓷上全髋关节置换术发生灾难性失败而未引发事件或外伤的情况。该患者是一名64岁的女性,患有变性性骨关节炎,并于3年前接受了全髋关节置换术。术中发现的内容包括完整的陶瓷股骨头,陶瓷衬套插入件略有损坏,弥漫性金属化病以及茎的耳轴过度磨损。在去除金属碎屑,切除类金属组织和大量灌洗关节后,将假体修改为模块化修改系统。尽管术前已对先前的手术报告进行了回顾,但没有迹象表明头颈部锥度不匹配。仅在进行翻修手术并引起高度怀疑之后,才对先前的植入物记录进行分析并确定不匹配。