摘要

Flow injection hydride generation atomic absorption spectrometry (FI-HGAAS) was used for determination of lead in dialysis concentrates.The parameters such as acidity,concentration of oxidising and reducing agents and argon gas flow rate were investigated to reach the best peak height sensitivity.No significant background signal was observed at high salt concentrations.The detection limit,concentration giving a signal equal to three times standard deviation of the blank signal,was 0.7 ng ml~(-1)' for a 500 mul injection volume.Precision of the measurements at the 20 ng ml~(-1) level was 3.7% R.S.D.The dialysis concentrates analysed by FI-HGAAS were found to have 10-70 ng ml~(-1) of lead.The same samples were analysed by ETAAS after removing the matrix using solid phase extraction with Chelex 100.The results were in agreement with those obtained by FI-HGAAS.
机译:用流动注射氢化物发生原子吸收光谱法(FI-HGAAS)测定透析浓缩液中的铅,研究了酸度,氧化剂和还原剂的浓度以及氩气流速等参数,以达到最佳的峰高灵敏度。在高盐浓度下观察到显着的背景信号。对于500 mul进样量,检测极限的浓度给出的信号等于空白信号标准偏差的三倍,为0.7 ng ml〜(-1)'。在20 ng ml〜(-1)浓度下的RSD为3.7%.FI-HGAAS分析的透析浓缩液中铅含量为10-70 ng ml〜(-1)。使用Chelex 100进行固相萃取得到的基质。结果与FI-HGAAS获得的结果一致。