Evaluation of luminol-H2O2-KIO4 chemiluminescence system and its application to hydrogen peroxide, glucose and ascorbic acid assays.

机译:鲁米诺-H2O2-KIO4化学发光系统的评估及其在过氧化氢,葡萄糖和抗坏血酸测定中的应用。

获取原文
获取原文并翻译 | 示例
           

摘要

Hydrogen peroxide can be directly determined by a luminol-KIO4-H2O2 CL system. The detection limit was 3.0 X 10-8 mol/litre and the calibration graph was linear over the range of 2.0X10-7-6.0X10-4 mol/litre. The relative standard deviation of H2O2 was 1.1% for 2.0X10-6 mol/litre (n=11). Glucose was indirectly determined through measuring the H2O2 generated by the oxidation of glucose in the presence of glucose oxidase at pH 7.6. The present method provides a source for H2O2, which, in turn, coupledwith the luminol-KIO4-H2O2 CL reaction system. The CL was linearly correlated with glucose concentration of 0.6-110μg/ml. The relative standard deviation was 2.1% for 10μg/ml (n=11). Detection limit of glucose was 0.08μg/ml. Ascorbic acid was also indirectly determined by the suppression of luminol-KIO4-H2O2 CL system. The calibration curve was linear over the range of 1.0X10-7-1.0X10-5 mol/litre of ascorbic acid. The relative standard deviation was 1.0% for 8.0X10-7 mol/litre (n=11). Detection limit of ascorbic acid was 6.0X10-8 mol/litre. These proposed methods have been applied to determine glucose and ascorbic acid in tablets and injection.
机译:过氧化氢可以通过鲁米诺-KIO4-H2O2 CL系统直接测定。检出限为3.0 X 10-8 mol / L,校准曲线在2.0X10-7-6.0X10-4 mol / L的范围内呈线性。对于2.0×10-6 mol / L(n = 11),H2O2的相对标准偏差为1.1%。通过在pH 7.6的葡萄糖氧化酶存在下测量葡萄糖氧化产生的H2O2间接测定葡萄糖。本方法提供了H 2 O 2的来源,其又与鲁米诺-KIO 4 -H 2 O 2 CL反应系统偶联。 CL与0.6-110μg/ ml的葡萄糖浓度线性相关。 10μg/ ml的相对标准偏差为2.1%(n = 11)。葡萄糖的检出限为0.08μg/ ml。抗坏血酸还可以通过抑制鲁米诺-KIO4-H2O2 CL体系间接测定。在1.0X10-7-1.0X10-5 mol / L抗坏血酸的范围内,校准曲线是线性的。 8.0X10-7 mol / L(n = 11)的相对标准偏差为1.0%。抗坏血酸的检出限为6.0×10-8 mol / L。这些建议的方法已应用于测定片剂和注射剂中的葡萄糖和抗坏血酸。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号