首页> 外文期刊>Talanta: The International Journal of Pure and Applied Analytical Chemistry >Uncertainty of ozone measurements with the primary standard reference photometer (SRP45)
【24h】

Uncertainty of ozone measurements with the primary standard reference photometer (SRP45)

机译:一级标准参比光度计(SRP45)的臭氧测量不确定度

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

A comparison of the ozone primary reference standard photometer serial number 45 (SRP45) against the National Institute of Standards and Technology (NIST) instruments, serial number 0 (SRP0) and 2 (SRP2), has been performed in order to establish the traceability and comparability of ozone measurements made by the Chilean atmospheric science community. A complete uncertainty budget was developed for SRP45, using a GUM approach. The results of the comparisons allow us to conclude that SRP45, SRP0 and SRP2 are comparable according to internationals criteria over an ozone mole fraction range of 0 nmol mol ~(-1) to at least 500 nmol mol ~(-1). The official result for the validation of SRP45 is x _(ozone) ~(SRP45)=0.013+0.99806x _(ozone) ~(SRP2)nmolmol ~(-1) with an expanded uncertainty of 2×(0.27) ~2+(1.18×10 ~(-2)×x _(ozone)) ~2 from 0 to 500 nmol mol ~(-1).
机译:为了确定可追溯性和可追溯性,将臭氧一级参考标准光度计序列号45(SRP45)与国家标准技术研究所(NIST)仪器序列号0(SRP0)和2(SRP2)进行了比较。智利大气科学界所做的臭氧测量结果的可比性。使用GUM方法为SRP45制定了完整的不确定性预算。比较的结果使我们得出结论,根据国际标准,SRP45,SRP0和SRP2在臭氧摩尔分数范围从0 nmol mol〜(-1)到至少500 nmol mol〜(-1)时是可比的。验证SRP45的官方结果是x _(臭氧)〜(SRP45)= 0.013 + 0.99806x _(臭氧)〜(SRP2)nmolmol〜(-1),扩展不确定度为2×(0.27)〜2 + (1.18×10〜(-2)×x _(臭氧))〜2从0到500 nmol mol〜(-1)。