摘要

Ozone level 1 calibration of ozone level 2 transfer standard with different ranges was carried out. The results showed that the slopes and the intercepts of the calibration curves were all within the ozone level 1 calibration index with the range of 500 nmol/mol and 1000 nmol/mol by ozone level 2 transfer standard. It was recommended to choose the right range as required when ozone level 2 transfer standard were calibrated. For the major monitoring sites, the range of 500 nmol/mol could meet the requirements. Furthermore, the ozone level 1 calibration period was observed with the same experiment conditions, which were serial communication port, 6 concentration points and the ozone transfer standard range of 500 nmol/mol. Within 36 months, the slopes were between 0.9927 and 1.0105 with the average slope 1.0019, and the intercepts were between -0.4373and -0.0266 nmol/mol with the average intercept -0.2115 nmol/mol, while the relative standard deviation (RSD) of the slopes was 0.32% and the standard deviation of the intercepts was 0.094 nmol/mol. The calibration results were all within the ozone level 1 calibration index. Consulting the relevant standards and the quality management requirements of ambient air ozone monitoring currently in China, it was suggested ozone level 2 transfer standard to perform ozone level 1 cali-bration once every 12 months at least.%对不同量程的臭氧二级传递标准进行一级校准实验,结果显示量程分别为500 nmol/mol和1000 nmol/mol时,所得校准曲线的斜率和截距均能够满足臭氧一级校准指标的要求.对臭氧二级传递标准开展臭氧一级校准时,应根据实际需求选择合适的量程.对于大多数的监测站点,臭氧二级传递标准量程范围选择为500 nmol/mol可满足要求.在相同实验条件下(选择串口连接模式、校准曲线浓度点个数为6个、臭氧二级传递标准量程为500 nmol/mol),考察臭氧一级校准的周期.36个月的考察结果显示臭氧一级校准曲线的斜率在0.9927~1.0105之间,平均斜率为1.0019;截距在-0.4373~-0.0266 nmol/mol之间,平均截距为-0.2115 nmol/mol;斜率的相对标准偏差为0.32%,截距的标准偏差为0.094 nmol/mol,均能符合臭氧一级校准指标的要求.结合国内外相关标准及我国环境空气臭氧监测质量管理需求,建议臭氧二级传递标准至少每12个月开展一次臭氧一级校准.

著录项

  • 来源
    《中国测试》 |2018年第1期|40-43|共4页
  • 作者

    王帅斌; 李宁; 樊强; 杜健;

  • 作者单位

    国家环境保护污染物计量和标准样品研究重点实验室 环境保护部标准样品研究所,北京 100029;

    国家环境保护污染物计量和标准样品研究重点实验室 环境保护部标准样品研究所,北京 100029;

    国家环境保护污染物计量和标准样品研究重点实验室 环境保护部标准样品研究所,北京 100029;

    国家环境保护污染物计量和标准样品研究重点实验室 环境保护部标准样品研究所,北京 100029;

  • 原文格式 PDF
  • 正文语种 chi
  • 中图分类
  • 关键词

    臭氧标准参考光度计; 臭氧二级传递标准; 臭氧一级校准; 量程; 校准周期;

相似文献

  • 中文文献
  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号