首页> 外文期刊>Urology >Mechanical reliability, surgical complications, and patient and partner satisfaction of the modern three-piece inflatable penile prosthesis.
【24h】

Mechanical reliability, surgical complications, and patient and partner satisfaction of the modern three-piece inflatable penile prosthesis.

机译:现代三件式可充气阴茎假体的机械可靠性,手术并发症以及患者和伙伴的满意度。

获取原文
获取原文并翻译 | 示例
       

摘要

OBJECTIVES: The modern three-piece inflatable penile prosthesis (IPP) has undergone multiple revisions since its introduction in 1973. We reviewed devices placed since the last major revision by American Medical Systems (AMS) in 1987. METHODS: A retrospective chart review was refined with data from an independent patient and partner survey. RESULTS: Two hundred twelve consecutive penile prosthetic devices placed by a single surgeon over an 8-year period are reviewed. One hundred sixty-nine of the devices were three-piece inflatables with 146 being primary implants. The average device has been in place 36.5 months (range 9 to 102). The infection rate in 146 primary three-piece devices was 2.1%. The infection rate in 46 secondary implants or revisions was 6.5%, excluding seven salvage attempts. Mechanical failure in 122 primary AMS devices placed was 4.1%. Mechanical failure in 24 Mentor devices was 4.2% if one discounts connector failures that were revised in 1990. A surgical complication and revision rate of 1.4% was noted in the 146 primary implants. An independent telephone survey achieved a 57% and 24% response rate in patients and partners with three-piece devices placed. In the group of 86 patients with a primary three-piece device placed and complete follow-up, the probability of having a normally functioning device placed in a single operative procedure was 90.6% at 3 years. On a 1 to 10 scale looking at all primary devices, secondary devices, revisions, and infections, the average and median satisfaction rate was as follows: 8.2, 8.5; 8.4, 9.0; 7.7, 7.75 for the Ultrex patients, CX 700 and Mentor patients, and all partners, respectively. CONCLUSIONS: The modern three-piece IPP is an excellent surgical option offering a very safe, reliable return to sexual activity for our patients.
机译:目的:现代的三件式可充气阴茎假体(IPP)自1973年推出以来已经历了多次修订。我们审查了自1987年美国医疗系统(AMS)的上次重大修订以来所放置的器械。方法:对回顾性图表进行了完善来自独立的患者和伴侣调查的数据。结果:回顾了由单个外科医生在8年内放置的212个连续的阴茎假体。其中的169个装置是三件式充气装置,其中146个是主要植入物。平均设备使用时间为36.5个月(范围为9到102)。 146个主要的三件式设备的感染率为2.1%。 46次二次植入或翻修的感染率为6.5%,其中不包括7次抢救尝试。在放置的122个主要AMS设备中,机械故障为4.1%。如果对1990年修订的连接器故障进行打折,则24个Mentor设备的机械故障率为4.2%。在146例初次植入物中,手术并发症和修订率为1.4%。一项独立的电话调查显示,放置三件式设备的患者及其伴侣的响应率分别为57%和24%。在86例患者中,放置了一个主要的三件式器械并进行了完整的随访,在3年时,一次手术中可以正常运行的器械的可能性为90.6%。以1到10的比例查看所有主要设备,次要设备,修订版和感染情况,平均满意度和中位数满意度分别为:8.2、8.5; 8.4、9.0; Ultrex患者,CX 700和Mentor患者以及所有伙伴分别为7.7、7.75。结论:现代的三件式IPP是一种出色的手术选择,可为我们的患者提供非常安全,可靠的性活动恢复。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号