首页> 外文期刊>Ultrasound in obstetrics & gynecology: the official journal of the International Society of Ultrasound in Obstetrics and Gynecology >Comparison of inter- and intraobserver agreement and reliability between three different types of placental volume measurement technique (XI VOCAL, VOCAL and multiplanar) and validity in the in-vitro setting.
【24h】

Comparison of inter- and intraobserver agreement and reliability between three different types of placental volume measurement technique (XI VOCAL, VOCAL and multiplanar) and validity in the in-vitro setting.

机译:三种不同类型的胎盘体积测量技术(XI VOCAL,VOCAL和多平面)之间的观察者之间和观察者之间的一致性和可靠性以及在体外环境中的有效性比较。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVES: To compare XI VOCAL (eXtended Imaging Virtual Organ Computer-aided AnaLysis) for three-dimensional (3D) ultrasound volumetry of the placenta and of phantom objects with a rotational method using VOCAL and with the multiplanar method. METHODS: We acquired 3D volume datasets from 32 fetuses at 11-14 weeks' gestation. Placental volume was calculated twice by each of two observers using XI VOCAL (with 5, 10, 15 and 20 slices), multiplanar (1-mm interval) and VOCAL (with 12 degrees, 18 degrees and 30 degrees rotation) methods. In addition, validity was assessed using the in-vitro setting with three phantom objects of known volume. RESULTS: Both inter- and intraobserver reliabilities were very high for all three methods. There was no systematic bias between any two methods except between XI VOCAL (10 slices) and the multiplanar (1-mm interval) method, with a smaller volume using the former method. The limits of agreement were wide between any two of the three methods. In the in-vitro setting, there was a trend towards less valid measurements with the XI VOCAL technique and fewer slices. With the same number of steps, measurements made with VOCAL (12 degrees and 18 degrees) were more valid than were those made with XI VOCAL (15 and 10 slices, respectively). CONCLUSION: XI VOCAL cannot be used interchangeably with VOCAL or multiplanar techniques in measuring placental volume at 11-14 weeks' gestation.
机译:目的:比较使用VOCAL的旋转法和多平面法对胎盘和幻影对象进行三维(3D)超声检查的XI VOCAL(扩展成像虚拟器官计算机辅助分析)。方法:我们在妊娠11-14周时从32胎中获得了3D体积数据集。两名胎盘观测者分别使用XI VOCAL(具有5、10、15和20个切片),多平面(间隔1mm)和VOCAL(具有12度,18度和30度旋转)方法计算两次胎盘体积。另外,使用体外装置对已知体积的三个体模对象评估有效性。结果:三种方法的观察者间和观察者内可靠性均很高。除了XI VOCAL(10片)和多平面(1-mm间隔)方法之间的任何两种方法之间,没有系统性的偏差,使用前一种方法的体积较小。三种方法中任何两种方法之间的协议限制都很大。在体外环境中,使用XI VOCAL技术进行有效测量的趋势越来越少,而且切片也越来越少。在相同的步骤数下,使用VOCAL(12度和18度)进行的测量比使用XI VOCAL(分别为15和10个切片)进行的测量更为有效。结论:XI VOCAL不能与VOCAL或多平面技术互换使用来测量妊娠11-14周的胎盘体积。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号