首页> 外文期刊>Ultrasound in Medicine and Biology >Evaluation of lens hardness in cataract surgery using high-frequency ultrasonic parameters in vitro.
【24h】

Evaluation of lens hardness in cataract surgery using high-frequency ultrasonic parameters in vitro.

机译:使用高频超声参数体外评估白内障手术中的晶状体硬度。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Ultrasonic parameters of sound velocity and frequency-dependent attenuation ranging from 25 to 45 MHz were measured for the purpose of evaluating the hardness of lenses in cataract surgery (phacoemulsification). Measurements were performed with a 35-MHz ultrasonic transducer on porcine lenses in which artificially cataracts were induced. The hardness of the cataractous lens was also evaluated by mechanical measurement of its elastic properties. The results indicated that the ultrasonic attenuation coefficients in normal porcine lenses were approximately 4.49 +/- 0.05 (mean +/- SD) and 6.32 +/- 0.04 dB/mm at 30 and 40 MHz, respectively. The development progression of the cataracts resulted in the attenuation coefficient increasing linearly to 7.36 +/- 0.25 and 11.1 +/- 0.92 dB/mm, respectively, corresponding to an increase of Young's modulus from 2.6 to 101.2 kPa. The sound velocity concomitantly increased from 1639.8 +/- 4.2 to 1735.6 +/- 10.4 m/s. Evaluation of the relationship between the phacoemulsification energy level and ultrasonic parameters in vitro by surgeons revealed that both the attenuation coefficient and sound velocity were linearly correlated with the phacoemulsification energy (r = 0.941 and 0.915, respectively). These results showed that measuring high-frequency ultrasonic parameters provides surgeons with good capability and reproducibility for selecting the optimal energy level for phacoemulsification.
机译:为了评估白内障手术(超声乳化)中晶状体的硬度,测量了超声速度的声速和随频率变化的衰减范围为25至45 MHz。用35-MHz超声换能器在人工晶状体被诱发的猪晶状体上进行测量。白内障晶状体的硬度也通过对其弹性性能的机械测量来评估。结果表明,普通猪晶状体在30 MHz和40 MHz下的超声衰减系数分别约为4.49 +/- 0.05(平均值+/- SD)和6.32 +/- 0.04 dB / mm。白内障的发展进程导致衰减系数分别线性增加至7.36 +/- 0.25和11.1 +/- 0.92 dB / mm,对应于杨氏模量从2.6增加到101.2 kPa。声速随之从1639.8 +/- 4.2提高到1735.6 +/- 10.4 m / s。外科医生对超声乳化能量水平与超声参数之间关系的评估表明,衰减系数和声速均与超声乳化能量线性相关(分别为r = 0.941和0.915)。这些结果表明,测量高频超声参数为外科医生提供了选择超声乳化最佳能量水平的良好能力和可重复性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号