Wire tension in the Ilizarov system: accuracy of the wire-tensioning device

机译:Ilizarov系统中的钢丝张力:钢丝张紧装置的精度

获取原文
获取原文并翻译 | 示例
           

摘要

The Ilizarov fixator consists of tensioned wires that attach bone segments to a modular frame. The aim of this study was to establish the accuracy and precision of the wire-tensioning device supplied with the Ilizarov external fixation system. The device was used to tension a wire in direct opposition to a calibrated load cell. Five subjects tested three devices, at each of their four tension settings, in two separate sessions. Subjects could not see the true tension during the test. There were significant differences between the results for different subjects (p<0.01) and instruments (p<0.01) but not for different tension settings or between the two sessions. Overall mean measured tensions were 4.9 per cent (standard deviation, 4.4 per cent) below intended values. Tensions obtained at the maximum edge (completely occluded) on the scale markings were significantly (p< 0.001) closer to the nominal values (mean discrepancy, 3.6 per cent) than those at the minimum edge (mean discrepancy, 17.6 per cent). Several factors influence wire tension. Tensioning devices are not identical and the results obtained with them depend on the user. If the scale markings are completely occluded, the discrepancy between intended and actual tensions of around 5 per cent is likely to be adequate for clinical practice since surgeons do not select the most suitable tension following quantitative data assessment, but rather it is a judgement based on surgical experience and consideration for the patient weight and expected level of activity.
机译:Ilizarov固定器由张紧的金属丝组成,这些金属丝将骨骼段连接到模块化框架上。这项研究的目的是确定随Ilizarov外固定系统一起提供的钢丝张紧装置的精度和精确度。该设备用于与校准的称重传感器直接相反地张紧电线。五名受试者在两个单独的环节中以四个张力设置中的每个测试了三个装置。在测试过程中,受试者看不到真正的紧张感。对于不同的受试者(p <0.01)和器械(p <0.01),结果之间存在显着差异,但对于不同的张力设置或两个疗程之间则没有差异。总体平均测得的张力比预期值低4.9%(标准偏差,4.4%)。在刻度尺标记的最大边缘处(完全被遮挡)获得的张力比最小边缘处的名义值(平均差异为3.6%)显着(p <0.001)更接近名义值(平均差异为3.6%)。有几个因素会影响钢丝的张力。张紧装置并不相同,使用它们所获得的结果取决于用户。如果刻度标记完全被遮挡,则由于临床医师在定量数据评估后并未选择最合适的张力,因此预期张力与实际张力之间的大约5%的差异可能足以满足临床实践的需要。手术经验以及对患者体重和预期活动水平的考虑。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号