首页> 外文期刊>Proceedings of the Institution of Mechanical Engineers, Part H. Journal of Engineering in Medicine >A finite element method comparison of wear in two metal-on-metal total hip prostheses
【24h】

A finite element method comparison of wear in two metal-on-metal total hip prostheses

机译:两种金属对金属全髋关节假体磨损的有限元方法比较

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The contact mechanics of two metal-on-metal (MOM) total hip prostheses was studied by means of the finite element method (FEM). The purpose of the work was to compare two total hip replacements (Durom and Metasul) with regard to the amount of wear debris released. Wear on the bearing surfaces was evaluated following Reye hypotheses from the pressure distribution, computed by means of three-dimensional FEM models; an approximate analytical model based on Hertz contact theory has also been developed and discussed. The results show that in the dry friction condition the Durom joint releases almost twice as much wear volume as produced by the Metasul joint. Therefore, while Durom implants can improve hip stability by increasing the prosthetic impingement-free range of motion (PIF-ROM), Metasul prostheses can be a valuable solution whenever wear represents a critical choice factor.
机译:通过有限元方法(FEM)研究了两种金属对金属(MOM)全髋关节假体的接触力学。这项工作的目的是比较两种全髋关节置换术(Durom和Metasul)的磨损碎片释放量。根据Reye假设,通过三维FEM模型计算得出的压力分布,评估轴承表面的磨损;还开发并讨论了基于赫兹接触理论的近似分析模型。结果表明,在干摩擦条件下,Durom接头释放的磨损量几乎是Metasul接头产生的磨损量的两倍。因此,尽管Durom植入物可通过增加无假体的运动范围(PIF-ROM)来改善髋关节稳定性,但只要磨损成为关键的选择因素,Metasul假体便是有价值的解决方案。