Traditional and the new IUPAC recommended calculation methods for determining the limit of detection and the limit of quantification are applied to the trace determination of cadmium and lead in water. The new method is based on the upper limit approach (ULA), which utilises the one-sided confidence band around the calibration line. The residual standard deviation of the calculated calibration line as well as the t-distribution critical value are used for the calculation of both limits. An appropriate design of the calibration experiment is described. The calculation of both limits in the new way is extremely easy when the calibration experiment is equidistantly designed.
机译:IUPAC推荐的传统和新的计算方法,用于确定检测限和定量限,用于水中痕量镉和铅的测定。新方法基于上限方法(ULA),该方法利用了校准线周围的单侧置信带。计算出的校准线的残留标准偏差以及t分布临界值用于计算两个极限。描述了校准实验的适当设计。等距设计校准实验时,以新的方式计算两个极限非常容易。