Portable Electrochemical Immunochromatography Platform for Point-of-care Testing and Comparison with Enzyme-linked Immunosorbent Assay

机译:便携式电化学免疫层析平台,进行护理点测试,与酶联免疫吸附测定的比较

获取原文
获取原文并翻译 | 示例
           

摘要

We conducted electrochemical immunochromatography using a developed electrochemical immunochromatography platform. We used immunoglobulin G (IgG) as a representative protein. We also conducted enzyme-linked immunosorbent assay (ELISA) as a control test. We used the same reagents except for the substrate and compared their limits of detection (LODs) and detection ranges. We succeeded in measuring IgG concentrations ranging from 30 ng/ml to 10 μg/ml and obtained 23.00 ng/ml as the LOD for electrochemical immunochromatography. The detection signal was not saturated even at 10 μg/ml, so it seems that higher concentrations of IgG can also be measured. The LOD was approximately 80-fold higher than that of ELISA (0.30 ng/ml); however, the detection range was one order of magnitude higher than that of ELISA. While electrochemical immunochromatography is less sensitive than ELISA, we demonstrated sufficient sensitivity for its use in a point-of-care testing (POCT) device. Furthermore, this information will be very helpful when introducing an assay system already established in ELISA to POCT.
机译:我们使用开发的电化学免疫层状平台进行电化学免疫层析。我们使用免疫球蛋白G(IgG)作为代表性蛋白质。我们还将酶联免疫吸附试验(ELISA)作为对照试验进行。除了基材外使用相同的试剂,并比较其检测限(LOD)和检测范围。我们成功地测量了从30ng / ml至10μg/ ml的IgG浓度,并获得23.00ng / ml作为电化学免疫层析的植物。即使在10μg/ ml下,检测信号也不饱和,因此似乎也可以测量较高浓度的IgG。 LOD大约高于ELISA(0.30ng / ml)的80倍;然而,检测范围比ELISA高出一个数量级。虽然电化学免疫层析比ELISA敏感,但我们对其在护理点测试(POCT)装置中的使用表现出足够的敏感性。此外,当在ELISA中建立的测定系统到POCT时,这些信息将非常有用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号