Performance and Evaluation of Smartphone Based Wireless Blood Pressure Monitoring System Using Bluetooth

机译:基于智能手机的蓝牙无线血压监测系统的性能与评估

获取原文
获取原文并翻译 | 示例
       

摘要

In medical practice, blood pressure is intensively studied parameter and often measured for diagnosis at primary healthcare level. In this paper, the state-of-the-art Bluetooth technology has been used for designing an innovative wireless blood pressure monitoring system to display the systolic and diastolic blood pressure values on Android-based smartphone. A pilot study was undertaken on 12 healthy subjects to validate the performance of developed prototype by comparing with OMRON HEM-7111 and mercury sphygmomanometer. Comparison shows an accuracy of 98.63% and 97.13% with OMRON HEM-7111 and 98.99% and 98% with mercury sphygmomanometer for systolic blood pressure and diastolic blood pressure, respectively. Bias value in Bland-Altman plots gives good level of agreement within 95% confidence limits. The developed prototype device mean and standard deviation of differences is less than 5 and 8 mmHg, respectively, and it fulfills the recommendation criteria as per the requirement of the Association for Advancement of Medical Instruments protocol.
机译:在医学实践中,对血压进行了深入研究,并经常在初级医疗保健水平对其进行测量以进行诊断。在本文中,最先进的蓝牙技术已用于设计创新的无线血压监测系统,以显示基于Android的智能手机的收缩压和舒张压值。通过与OMRON HEM-7111和水银血压计进行比较,对12个健康受试者进行了初步研究,以验证所开发原型的性能。比较显示,使用OMRON HEM-7111的准确度分别为98.63%和97.13%,使用汞血压计的准确度分别为98.99%和98%。在Bland-Altman图中的偏差值在95%的置信范围内提供了良好的一致性。所开发的原型设备的均值和标准差的差异分别小于5和8 mmHg,并且符合医疗器械进步协会协议的推荐标准。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号