Instrument compliance effects revisited: linear viscoelastic measurements

机译:再谈仪器的顺应性效应:线性粘弹性测量

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Torsional compliance of the torque transducer can be an important issue in linear viscoelastic (LVE) measurements when the sample stiffness is high relative to the instrument stiffness. We evaluated compliance effects of the ARES 2K-FRT on LVE measurements by systematically comparing the results of the frequency sweep mode obtained with 25 and 8 mm plates, respectively. In addition to the transducer, the test fixtures do contribute significantly to the system compliance. Without correction, the upper limit for the complex modulus ∣ G ∗ ∣ is approximately close to 4 × 105 Pa at 10% uncertainty, when using 25 mm plates. This limit is lower than the plateau modulus $left( {G_N^0 } right)$ of most polymers. Therefore, instrument compliance can lead to significant errors for $G_N^0 $ and wrong scaling for G ″ in the plateau and Rouse regions. The respective roles of transducer and tool compliances are discussed. The FRT transducer compliance is corrected in real time in the instrument firmware. Tool compliance is a common problem for all rheometers when measuring stiff samples.
机译:当样品刚度相对于仪器刚度较高时,扭矩传感器的扭转柔度可能是线性粘弹性(LVE)测量中的重要问题。我们通过系统地比较分别用25和8 mm平板获得的扫频模式的结果,评估了ARES 2K-FRT对LVE测量的顺应性影响。除换能器外,测试夹具确实对系统合规性做出了重要贡献。如果不进行校正,则在使用25 mm钢板时,复模量∣ G ∗ ssup> upper的上限在10%不确定度下大约接近4×105 Pa。该极限低于大多数聚合物的平稳模量$ left({G_N ^ 0} right)$。因此,在高原地区和鲁斯地区,仪器的合规性可能导致$ G_N ^ 0 $的严重错误以及G''的错误缩放。讨论了换能器和工具合规性的各自作用。仪器固件中实时校正了FRT传感器的兼容性。对于所有流变仪,在测量硬质样品时,工具的合规性是一个常见问题。

著录项

  • 来源
    《Rheologica Acta》 |2011年第6期|p.537-546|共10页
  • 作者单位

    Beijing National Laboratory for Molecular Sciences, CAS Key Laboratory of Engineering Plastics, Joint Laboratory of Polymer Science and Materials, Institute of Chemistry, Chinese Academy of Sciences, Beijing, 100190, China;

    TA Instruments, 109 Lukens Drive, New Castle, DE, 19720, USA;

    TA Instruments, 109 Lukens Drive, New Castle, DE, 19720, USA;

    TA Instruments, Helfmann Park 10, 65760, Eschborn, Germany;

    Unité de Chimie et de Physique des Hauts Polymères, Université catholique de Louvain, 1348, Louvain-La-Neuve, Belgium;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    Compliance; Dynamic moduli; Viscoelasticity; Rheometer;

    机译:柔度;动态模量;粘弹性;流变仪;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号