A new technique of x-ray focusing crystal spectrometers’ calibration is the desired result. For this purpose the spectrometer is designed to register radiated copper Kα and Kβ lines by using a flat α-quartz crystal. This experiment uses pre-breakdown x-ray emissions in low vacuum of about 2.5–3 mbar. At this pressure the pinch will not form so the plasma will not radiate. The anode material is copper and the capacity of the capacitor bank is 22.6 nF. This experiment designed and mounted a repetitive triggering system to save the operator time making hundreds of shots. This emission amount is good for calibration and geometrical adjustment of an optical crystal x-ray focusing spectrometer.
机译:期望的结果是采用X射线聚焦晶体光谱仪进行校准的新技术。为此,该光谱仪设计为通过使用扁平的α石英晶体来配准辐射的铜Kα和Kβ线。该实验使用了约2.5–3 mbar的低真空下的分解前X射线发射。在此压力下,将不会形成收缩,因此等离子体将不会辐射。阳极材料为铜,电容器组的容量为22.6 nF。该实验设计并安装了重复触发系统,以节省操作员数百张照片的时间。该发射量有利于光学晶体X射线聚焦光谱仪的校准和几何调整。