首页> 外文期刊>Radiation Protection Dosimetry >OPTICAL IMAGES OF DOSE DISTRIBUTIONS IN GEL-FRICKE: DOSIMETRIC PERFORMANCES OF THE GEL
【24h】

OPTICAL IMAGES OF DOSE DISTRIBUTIONS IN GEL-FRICKE: DOSIMETRIC PERFORMANCES OF THE GEL

机译:凝胶弗里克斯剂量分布的光学图像:凝胶的定量性能

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The purpose of this work was to examine the dosimetric performances of the radiochromic Fricke-Agarose-Xylenol Orange gel by optical measurements in order to perform dose reconstructions, in view of a future development for 3-D maps. Optical images and dose-response curves of the gel were obtained by a CCD-based device, originally designed for reading radiochromic films, that was modified to meet the optical properties of the dosemeter. With a resolution of 0.18 x 0.18 mm the optimum range of doses in which per cent uncertainty is lower than 2% was 3-10 Gy. The minimum detectable dose, estimated as the absorbed dose corresponding to 3 SD above background, was 0.1 Gy. With a resolution of 1.98 x 1.98 mm the optimum range of doses in which per cent uncertainty is lower than 2% was 0.3-10 Gy. The minimum detectable dose, estimated as the absorbed dose corresponding to 3 SD above background, was 0.015 Gy. The comparison with alanine dosemeters in the dose range 7-10 Gy showed agreement within a few per cent and the same agreement was observed for the comparison with TLD in the range 1-3 Gy.
机译:鉴于3-D图的未来发展,这项工作的目的是通过光学测量检查放射变色的Fricke-琼脂糖-二甲苯酚橙凝胶的剂量学性能,以便进行剂量重建。凝胶的光学图像和剂量响应曲线是通过基于CCD的设备获得的,该设备最初被设计用于读取放射色膜,该设备经过修改以满足剂量计的光学特性。分辨率为0.18 x 0.18 mm,不确定性百分比低于2%的最佳剂量范围是3-10 Gy。最小可检测剂量为0.1 Gy,估计为对应于本底以上3 SD的吸收剂量。分辨率为1.98 x 1.98 mm,不确定度低于2%的最佳剂量范围为0.3-10 Gy。最低可检测剂量估计为对应于背景以上3 SD的吸收剂量,为0.015 Gy。与剂量范围为7至10 Gy的丙氨酸剂量计的比较表明一致性在百分之几以内,与TLD范围为1-3 Gy的比较也观察到相同的一致性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号