Fluoroscopy quality assurance measurements: automatic exposure rate control and image quality

机译:荧光检查质量保证测量:自动曝光率控制和图像质量

获取原文
获取原文并翻译 | 示例
           

摘要

A fluoroscopy system is a relatively complicated piece of equipment mad eup of several sub-components. For the operator to achieve good image quality without the overhead of high patient and staff doses, it is essential that all of these components are set up correctly. The automatic exposure rate control (AEC) and image quality are important aspects of a fluoroscopy system. Two principal parameters are used to assess AEC function: air kerma rate at the input phosphor with 1.0 mm copper filtration added and entrance surface dose rate for a 20 cm Perspex phantom.
机译:荧光检查系统是由几个子组件组成的相对复杂的设备。为了使操作员获得良好的图像质量而又不增加高额的患者和医务人员的开销,必须正确设置所有这些组件。自动曝光率控制(AEC)和图像质量是荧光检查系统的重要方面。两个主要参数用于评估AEC功能:输入磷光体的空气比释动能速率(添加1.0毫米铜过滤)和20厘米有机玻璃幻象的入射表面剂量率。