The investigations deal with the search for a passive dosemeter applicable to patient dosimetry in diagnostic radi- ology which meets particularly the requirements of a low detection limit and an adequate energy response, as well as sufficiently high precision and reliability. A thermoluminescence dosimetry (TLD) technology is reported that fits there requirements. The detector material consists of lithim fluoride co-doped with copper (LiF:Mg,Cu,P). The detectors, originating form China, are shown to exhibit high radiation sensitivity exceeding that of conventional lithium fluoride, LiF:Mg,Ti(TLD-100), by far. Based on the new technology a detection threshold of 0.1μGy, precision as low as<=1/100 (1σ_rel) and strict linearity of the TL signal- dose relationship were found, for doses up to at least 1 mGy.
机译:该调查涉及寻找适用于诊断放射学中患者剂量测定的无源剂量计,该无源剂量计特别满足低检测限和足够的能量响应以及足够高的精度和可靠性的要求。据报道,热发光剂量法(TLD)技术可以满足这些要求。检测器材料由氟化锂与铜共掺杂(LiF:Mg,Cu,P)组成。源自中国的探测器显示出远远超过常规氟化锂LiF:Mg,Ti(TLD-100)的辐射灵敏度。基于新技术,对于至少1 mGy的剂量,发现了0.1μGy的检测阈值,低至<= 1/100(1σ_rel)的精确度和TL信号-剂量关系的严格线性。