USE OF THE DICHROMATE SOLUTION AS A DOSIMETER FOR HIGH DOSE AND HIGH DOSE RATE

机译:将二氯酸盐溶液用作剂量计以实现高剂量和高剂量率

获取原文
获取原文并翻译 | 示例
       

摘要

The radiolytic reduction of Cr(Ⅵ) in N_2O-saturated aqueous solutions has been investigated over a large range of pH by steady state radiolysis. At pH 3 in the presence of formate as scavenger for •OH, G( — Cr(Ⅵ)) is higher than the expected yield and a strong dose rate effect is observed. The solution cannot be used for dosimetry at this pH. The proposed mechanism to explain the enhanced response and the dose rate dependence involves the reduction of the unstable radiolytically produced Cr(Ⅳ) by HCO_2~-. At pH 9.2 the reaction between Cr(Ⅳ) and formate is absent and no dose rate effect was observed up to 70 kGy min~(-1), which makes this system particularly attractive as a dosimeter in the 0.1-10 kGy dose range. Radiation chemical aspects relating to the use of this system as a high-dose and high-dose rate chemical dosimeter are discussed.
机译:通过稳态辐射分解研究了在大pH范围内N_2O饱和水溶液中Cr(Ⅵ)的辐射还原。在pH为3的条件下,在甲酸作为清除剂•OH的情况下,G(-Cr(Ⅵ))高于预期收率,并且观察到了强烈的剂量率效应。该溶液不能在此pH值下用于剂量测定。所提出的解释增强的反应和剂量率依赖性的机制涉及通过HCO_2〜-减少不稳定的放射分解产生的Cr(Ⅳ)。在pH 9.2时,不存在Cr(Ⅳ)与甲酸酯之间的反应,直至70 kGy min〜(-1)都没有观察到剂量率效应,这使得该系统作为剂量计在0.1-10 kGy剂量范围内特别有吸引力。讨论了与使用该系统作为高剂量和高剂量率化学剂量计有关的辐射化学方面。