首页> 外文期刊>Radiation Physics and Chemistry >TEMPERATURE, HUMIDITY AND TIME. COMBINED EFFECTS ON RADIOCHROMIC FILM DOSIMETERS
【24h】

TEMPERATURE, HUMIDITY AND TIME. COMBINED EFFECTS ON RADIOCHROMIC FILM DOSIMETERS

机译:温度,湿度和时间。对放射性铬胶片剂量仪的综合作用

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

The effects of both relative humidity and temperature during irradiation on the dose response of FWT-60-00 and Ris0 B3 radiochromic film dosimeters have been investigated in the relative humidity (RH) range 11-94% and temperature range 20-60℃ for irradiation by ~(60)Co photons and 10-MeV electrons. The results show that humidity and temperature cannot be treated as independent variables, rather there appears to be interdependence between absorbed dose, temperature, and humidity. Dose rate does not seem to play a significant role. The dependence of temperature during irradiation is + 0.25 ±0.1% per ℃ for the FWT-60-00 dosimeters and +0.5±0.1% per ℃ for Riso B3 dosimeters at temperatures between 20 and 50℃ and at relative humidities between 20 and 53%. At extreme conditions both with respect to temperature and to humidity, the dosimeters show much stronger dependences. Whenever possible one should use dosimeters sealed in pouches under controlled intermediate humidity conditions (30-50%) or, if that is impractical, one should maintain conditions of calibration as close as possible to the conditions of use. Without that precaution, severe dosimetry errors may result.
机译:在相对湿度(RH)范围为11-94%和温度范围为20-60℃的条件下,研究了辐照过程中相对湿度和温度对FWT-60-00和Ris0 B3放射致变色膜剂量计剂量响应的影响。由〜(60)Co光子和10-MeV电子组成。结果表明,湿度和温度不能视为独立变量,而是吸收剂量,温度和湿度之间似乎存在相互依存关系。剂量率似乎并未发挥重要作用。对于FWT-60-00剂量计,辐照过程中温度的依赖性为+ 0.25±0.1%/℃,对于Riso B3剂量计,在20至50℃和相对湿度为20%至53%时,温度的依赖性为+ 0.5±0.1%/℃ 。在关于温度和湿度的极端条件下,剂量计显示出更强的依赖性。只要有可能,就应该在受控的中等湿度条件下(30-50%)使用密封在小袋中的剂量计,或者,如果不切实际,应保持校准条件尽可能接近使用条件。否则,可能会导致严重的剂量学错误。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号