Dose-Rate and Humidity Effects Upon the gamma-Radiation Response of Nylon-Based Radiachromic Film Dosimeters

机译:剂量率和湿度对尼龙基Radiachromic薄膜剂量计的γ辐射响应的影响

获取原文

摘要

At dose-rates typical for exp 60 Co gamma irradiation sources, the radiation response of hexahydroxyethyl pararosaniline cyanide/ 50 mu m nylon radiachromic films is dependent upon dose-rate as well as upon the moisture content of the films, or the relative humidity of the surrounding atmosphere, respectively. Under equilibrium moisture conditions, the response measured at 606 nm 24 hours after end of irradiation shows its highest dose-rate dependence at about 32 % r.h. A decrease in dose-rate from 2.8 to 0.039 Gy.s exp -1 results in a decrease in response by 17%. At higher humidities, the sensitivity of the film as well as the rate dependence decreases and at 86% r.h. no discernible dose-rate effect could be found. At lower humidities than 32% a flat maximum in response follows. At nominal 0% r.h. a second absorption band at 412 nm appears which is converted completely to an additional 606 nm absorption by exposure to a humid atmosphere. After that procedure the resultant response is somewhat lower than but shows almost the same dose-rate dependence as at 32% r.h. or else to eliminate the dose-rate effect by an extrapolation procedure based on the fact that the rate dependence vanishes at zero dose. (Atomindex citation 11:547029)