首页> 外文期刊>Nuclear Instruments & Methods in Physics Research >MR-guided data framing for PET motion correction in simultaneous MR-PET: A preliminary evaluation
【24h】

MR-guided data framing for PET motion correction in simultaneous MR-PET: A preliminary evaluation

机译:MR引导的数据成帧以同时进行MR-PET进行PET运动校正:初步评估

获取原文
获取原文并翻译 | 示例
           

摘要

Head motion can significantly degrade image quality of static and dynamic Positron Emission Tomography (PET) of the human brain. One method to regain acceptable image quality in the presence of motion is to include the correction for motion in the reconstruction process. When applying motion correction, the PET data can be segmented into discrete parts of similar head position, referred to as frames. This framing of the data can reduce the computational overhead necessary for motion correction during the reconstruction process by reducing the number of discrete head positions which have to be accounted for. Here a framing algorithm is presented which minimises residual motion in the framed data, while taking full advantage of the additional information provided by Magnetic Resonance Imaging (MRI) in a simultaneous MR-PET acquisition. In the work presented here information on motion is derived from EPI sequences acquired simultaneously with the PET data. A comparison to images reconstructed with regular framing show a more clearly delineated cortex due to increased contrast between grey matter and white matter. This improvement in image quality is achieved as well as a reduction in the number of frames, thereby reducing the reconstruction time. Preliminary data indicates an efficient reduction of residual intra-frame motion compared to regular framing.
机译:头部运动会严重降低人脑静态和动态正电子发射断层扫描(PET)的图像质量。在存在运动时恢复可接受的图像质量的一种方法是在重建过程中包括对运动的校正。在进行运动校正时,可以将PET数据分割为具有相似头部位置的离散部分,称为帧。数据的这种成帧可以通过减少必须考虑的离散磁头位置的数量来减少重建过程中运动校正所需的计算开销。这里提出了一种成帧算法,该算法将帧数据中的残留运动最小化,同时在同步MR-PET采集中充分利用了磁共振成像(MRI)提供的附加信息。在此处介绍的工作中,有关运动的信息是从与PET数据同时获取的EPI序列中得出的。与通过规则取景重建的图像相比,由于灰质和白质之间的对比度增加,皮层轮廓更加清晰。实现了图像质量的改善以及帧数的减少,从而减少了重建时间。初步数据表明,与常规取景相比,有效减少了残留帧内运动。

著录项

  • 来源
  • 作者单位

    Institute of Neuroscience and Medicine (INM-4), Forschurxgszentrum Juelch GmbH, Germany;

    Institute of Neuroscience and Medicine (INM-4), Forschurxgszentrum Juelch GmbH, Germany;

    Institute of Neuroscience and Medicine (INM-4), Forschurxgszentrum Juelch GmbH, Germany;

    Institute of Neuroscience and Medicine (INM-4), Forschurxgszentrum Juelch GmbH, Germany;

    Institute of Neuroscience and Medicine (INM-4), Forschurxgszentrum Juelch GmbH, Germany;

    Institute of Neuroscience and Medicine (INM-4), Forschurxgszentrum Juelch GmbH, Germany;

    Institute of Neuroscience and Medicine (INM-4), Forschurxgszentrum Juelch GmbH, Germany;

    Institute of Neuroscience and Medicine (INM-4), Forschurxgszentrum Juelch GmbH, Germany;

    Institute of Neuroscience and Medicine (INM-4), Forschurxgszentrum Juelch GmbH, Germany;

  • 收录信息
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    positron emission tomography (PET); hybrid mr/pet scanner; motion correction; data framing;