首页> 外文会议>IEEE International Symposium on Biomedical Imaging >MR-based attenuation map re-alignment and motion correction in simultaneous brain MR-PET imaging
【24h】

MR-based attenuation map re-alignment and motion correction in simultaneous brain MR-PET imaging

机译:同时进行脑MR-PET成像的基于MR的衰减图重新对准和运动校正

获取原文

摘要

Head movement is a major issue in dynamic PET imaging. A simultaneous MR-PET scanner is capable of acquiring both MR and PET data concurrently, which enables opportunities to use MR information for PET motion correction. Here we propose an MR-based method to detect head motion and to correct motion artefacts during PET image reconstruction. The method is based on co-registration of multiple MR contrasts to extract motion parameters. The motion parameters are then used to guide the Multiple Acquisition Frame (MAF) algorithm to bin the PET list-mode data into multiple frames whenever significant motion occurs. Furthermore, motion parameters are used to re-align the PET attenuation u-map to each frame prior to the image reconstruction. Finally, PET images are reconstructed for each frame and combined to produce a final image. Using both phantom and in-vivo human data, we show that this method can significantly increase image quality and reduce motion artefacts.
机译:头部运动是动态PET成像中的主要问题。同时MR-PET扫描仪能够同时获取MR和PET数据,这使您有机会将MR信息用于PET运动校正。在这里,我们提出了一种基于MR的方法来检测头部运动并纠正PET图像重建过程中的运动伪影。该方法基于多个MR对比的共配准以提取运动参数。然后,只要出现重大运动,运动参数就可以用来指导多采集帧(MAF)算法,将PET列表模式数据分为多个帧。此外,运动参数用于在图像重建之前将PET衰减u-map重新对准每一帧。最终,为每帧重建PET图像并合并以生成最终图像。使用幻像和体内人类数据,我们表明该方法可以显着提高图像质量并减少运动伪像。