Scanning electron microscopy study of cavity preparation in deciduous teeth using the Er:YAG laser with different powers

机译:使用不同功率的Er:YAG激光的扫描电镜研究乳牙蛀牙的制备

获取原文
获取原文并翻译 | 示例
           

摘要

Using scanning electron microscopy (SEM) we evaluated the morphology of cavity surfaces in deciduous teeth prepared in vitro with the Er:YAG laser with different power parameters. Eight extracted cavity-free deciduous teeth with an intact crown were prepared using a traditional handpiece or an Er:YAG laser with different parameters (10 Hz/200 mJ, 10 Hz/300 mJ and 10 Hz/400 mJ). Samples were then processed and cavity surface morphology was evaluated by SEM to detect open dentinal tubules, or melting or cracking of the dentin. SEM showed that laser cavity preparation in deciduous teeth using different parameters left no smear layer and the dentinal tubules were clear. Dentin melting was not seen after cavity preparation at 200 mJ or 300 mJ, while visible dentin melting and cracks were detected at 400 mJ. The use of the laser at 10 Hz/200 mJ and 10 Hz/300 mJ for cavity preparation in deciduous teeth is safe and effective, but higher powers may damage the dentin.
机译:使用扫描电子显微镜(SEM),我们评估了用Er:YAG激光体外制备的具有不同功率参数的乳牙腔表面的形态。使用传统的机头或具有不同参数(10 Hz / 200 mJ,10 Hz / 300 mJ和10 Hz / 400 mJ)的Er:YAG激光,准备了八颗具有完整冠的无腔乳牙。然后处理样品,并通过SEM评估腔体表面形态,以检测开放的牙本质小管或牙本质的融化或破裂。扫描电镜显示,使用不同的参数在乳牙中进行激光腔准备没有留下涂片层,并且牙本质小管是透明的。在200兆焦耳或300兆焦耳的腔体准备后,未见牙本质融化,而在400兆焦耳处可见可见的牙本质融化和裂缝。使用10 Hz / 200 mJ和10 Hz / 300 mJ的激光在乳牙中准备蛀牙是安全有效的,但是更高的功率可能会损坏牙本质。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号