The thermal to epithermal neutron flux ratio (f) and the deviation of the epithermal neutron spectrum from the 1/E shape (α) are essential parameters for the correct application of k 0-standardized neutron activation analysis. Several methods are applied for the determination of f and α. They are based on Cd-covered multi-monitor or on bare-irradiations methods. The recently developed and characterized synthetic multi-element standards (SMELS) were designed as a validation tool for the proper implementation of the k 0-NAA method in a laboratory. In particular, SMELS Type III contains Au and Zr, thus allowing the direct determination of f and α. It could, therefore, replace the traditional flux monitors. Furthermore, it could be used as a quality control material to monitor the stability of the irradiation facility and the detector. This paper presents the accuracy of the f and α determination and the feasibility of quality control using SMELS for irradiation channel Y4 of the BR1 reactor.
机译:热与超热中子通量之比(f)和超热中子光谱与1 / E形状的偏离(α)是正确应用k 0 标准化中子活化分析的必要参数。几种方法可用于确定f和α。它们基于Cd覆盖的多显示器或裸照射方法。最近开发和表征的合成多元素标准(SMELS)被设计为验证工具,用于在实验室中正确实施k 0 -NAA方法。特别是,SMELS III型包含Au和Zr,因此可以直接确定f和α。因此,它可以代替传统的流量监控器。此外,它可用作质量控制材料,以监控照射设备和检测器的稳定性。本文介绍了BR1反应堆辐射通道Y4的f和α测定的准确性以及使用SMELS进行质量控制的可行性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号