【24h】

Quality assurance of an image guided intracranial stereotactic positioning system for radiosurgery treatment with helical tomotherapy

机译:影像引导颅内立体定位系统的质量保证,用于螺旋线放射疗法的放射外科治疗

获取原文
获取原文并翻译 | 示例
           

摘要

The aim of this work was to determine the accuracy and precision of stereotactic localization and treatment delivery using a helical tomotherapy based stereotactic radiosurgery (SRS) system. A tomotherapy specific radiosurgery workflow was designed that exploits the system’s on board megavotage CT (MVCT) imaging system so that it not only provides a pre-treatment volumetric verification image that can be used for stereotactic localization, eliminating the need for a patient-frame based coordinate system, but also supplies the treatment planning image. Using an imaging guidance based intracranial stereotactic positioning system, a head ring and tabletop docking device are used only for fixation, while image guidance is used for localization. Due to the unconventional workflow, a methodology for determining the localization accuracy was developed and results were compared to other linear accelerator based radiosurgery systems. In this work, the localization error using volumetric localization was found to be 0.45 mm ± 0.17 mm, indicating a localization precision of 0.3 mm within a 95% confidence interval. In addition, procedures for testing the delivery accuracy of the Tomotherapy system are described. Results show that the accuracy of the delivery can be verified to within ±1 voxel dimension. These results are well within conventional SRS tolerances and compare favorably to other linear accelerator based techniques.
机译:这项工作的目的是确定使用基于螺旋断层扫描的立体定向放射外科(SRS)系统确定立体定向和治疗的准确性和精确度。设计了特定于放射疗法的放射外科工作流程,该工作流程利用了系统的车载兆伏CT(MVCT)成像系统,因此它不仅提供了可用于立体定位的治疗前体积验证图像,而无需基于患者框架坐标系统,还提供治疗计划图像。使用基于成像引导的颅内立体定位系统,头环和桌面对接设备仅用于固定,而图像引导用于定位。由于采用了非常规的工作流程,因此开发了一种确定定位精度的方法,并将结果与​​其他基于线性加速器的放射外科系统进行了比较。在这项工作中,发现使用体积定位的定位误差为0.45 mm±0.17 mm,表明在95%置信区间内的定位精度为0.3 mm。此外,还介绍了测试Tomotherapy系统递送准确性的程序。结果表明,可以在±1体素尺寸范围内验证传递的准确性。这些结果完全在常规SRS容差范围内,并且可与其他基于线性加速器的技术进行比较。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号