金属フィルタを用いた循環器専用撮影装置における放射線被ばく低減の検討

机译:金属过滤器减少心血管成像系统辐射的研究

获取原文
获取原文并翻译 | 示例
           

摘要

In interventional X-ray for cardiology of flat panel digital detector (FPD), the phenomenon that exposure dose was suddenly increased when a subject thickness was thickened was recognized. At that time, variable metal built-in filters in FPD were all off. Therefore, we examined whether dose reduction was possible without affecting a clinical image using metal filter (filter) which we have been conventionally using for dose reduction. About 45% dose reduction was achieved when we measured an exposure dose at 30 cm of acrylic thickness in the presence of a filter. In addition, we measured signal to noise ratio/contrast to noise ratio/a resolution limit by the visual evaluation, and there was no influence by filter usage. In the clinical examination, visual evaluation of image quality of coronary angiography (40 cases) using a 5-point evaluation scale by a physician was performed. As a result, filter usage did not influence the image quality (p=NS). Therefore, reduction of sudden increase of exposure dose was achieved without influencing an image quality by adding filter to FPD.%可変金属フィルタを搭載しているFPD装置においrnて,追加フィルタを装着することによって厚い被写体厚rnによる急激な被ばく線量増加を一定量抑制することがrn可能となった.
机译:在平板数字检测器(FPD)的心脏病介入X射线中,发现了当对象厚度增加时曝光剂量突然增加的现象。当时,FPD中的可变金属内置过滤器全部关闭。因此,我们检查了是否可以使用金属过滤器(过滤器)减少剂量而不影响临床图像,而金属过滤器是我们传统上用于减少剂量的过滤器。当我们在有过滤器的情况下,在丙烯酸厚度为30 cm的地方测量暴露剂量时,可减少约45%的剂量。另外,我们通过视觉评估测量了信噪比/对比度/噪声比/分辨率极限,并且没有受到滤波器使用的影响。在临床检查中,由医师使用5分评估量表对冠状动脉造影(40例)的图像质量进行了视觉评估。结果,过滤器的使用不影响图像质量(p = NS)。因此,通过在FPD上添加滤镜,可以在不影响图像质量的情况下减少曝光剂量的突然增加。な被ばく线量増加を一定量抑制することがrn可能となった。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号