The determination of the efficiency of a Compton suppressed HPGe detector using Monte Carlo simulations

机译:使用蒙特卡洛模拟确定康普顿抑制HPGe探测器的效率

获取原文
获取原文并翻译 | 示例
       

摘要

A Compton suppressed high-purity germanium (HPGe) detector is well suited to the analysis of low levels of radioactivity in environmental samples. The difference in geometry, density and composition of environmental calibration standards (e.g. soil) can contribute to excessive experimental uncertainty to the measured efficiency curve. Furthermore multiple detectors, like those used in a Compton suppressed system, can add complexities to the calibration process. Monte Carlo simulations can be a powerful complement in calibrating these types of detector systems, provided enough physical information on the system is known. A full detector model using the Geant4 simulation toolkit is presented and the system is modelled in both the suppressed and unsuppressed mode of operation. The full energy peak efficiencies of radionuclides from a standard source sample is calculated and compared to experimental measurements. The experimental results agree relatively well with the simulated values (within ~5 - 20%). The simulations show that coincidence losses in the Compton suppression system can cause radionuclide specific effects on the detector efficiency, especially in the Compton suppressed mode of the detector. Additionally since low energy photons are more sensitive to small inaccuracies in the computational detector model than high energy photons, large discrepancies may occur at energies lower than ~ 100 keV.
机译:康普顿抑制的高纯锗(HPGe)检测器非常适合分析环境样品中的低放射性。环境校准标准品(例如土壤)的几何形状,密度和成分的差异可能导致测量效率曲线的实验不确定性过大。此外,像康普顿抑制系统中使用的检测器一样,多个检测器可能会增加校准过程的复杂性。如果已知有关系统的足够物理信息,则蒙特卡洛模拟可以很好地校准这些类型的检测器系统。提出了使用Geant4仿真工具包的完整检测器模型,并在抑制和未抑制操作模式下对系统进行了建模。计算来自标准来源样品的放射性核素的全能峰效率,并将其与实验测量值进行比较。实验结果与模拟值比较吻合(在〜5-20%以内)。仿真表明,康普顿抑制系统中的重合损耗会导致放射性核素对探测器效率的特定影响,尤其是在康普顿抑制模式下。另外,由于低能光子比高能光子对计算探测器模型中的小误差更敏感,因此在能量低于100 keV时可能会出现较大差异。

著录项

  • 来源
    《Journal of Environmental Radioactivity》 |2012年第2012期|p.1-7|共7页
  • 作者单位

    Australian Nuclear Science and Technology Organisation (ANSTO), Lucas Heights, NSW Australia;

    Australian Nuclear Science and Technology Organisation (ANSTO), Lucas Heights, NSW Australia;

    Australian Nuclear Science and Technology Organisation (ANSTO), Lucas Heights, NSW Australia;

    Australian Nuclear Science and Technology Organisation (ANSTO), Lucas Heights, NSW Australia;

  • 收录信息 美国《科学引文索引》(SCI);美国《工程索引》(EI);美国《生物学医学文摘》(MEDLINE);美国《化学文摘》(CA);
  • 原文格式 PDF
  • 正文语种 eng
  • 中图分类
  • 关键词

    HPGe detector; efficiency calibration; monte carlo; environmental sample;

    机译:HPGe检测器;效率校准;蒙特卡洛;环境样品;

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号