首页> 外文期刊>Journal of Clinical Monitoring and Computing >Low minimum alveolar concentration alarm: a standard for prevention of awareness during general anaesthesia maintained by inhalational anaesthetics
【24h】

Low minimum alveolar concentration alarm: a standard for prevention of awareness during general anaesthesia maintained by inhalational anaesthetics

机译:最低肺泡浓度低警报:通过吸入麻醉药预防全身麻醉期间意识的标准

获取原文
获取原文并翻译 | 示例

摘要

Awareness during general anaesthesia is a rare but significant problem that can be frightening to the patients. We suggest that newer generation monitors should include this facility to provide a low alarm limit to MAC settings so as to improve the quality of patient care. Also we suggest that a “near empty” alarm be incorporated into vaporizers which can warn the anaesthesiologist prior to development of possible light plane of anaesthesia. We hope that adopting these two features can help enhance patient safety and can further aid in quality assurance.
机译:全身麻醉期间的意识是一个罕见但重要的问题,可能使患者感到恐惧。我们建议新一代监护仪应包括此功能,以降低对MAC设置的警报限制,从而提高患者护理质量。我们还建议将“接近空”警报结合到汽化器中,以便在可能的麻醉光平面发展之前向麻醉师发出警告。我们希望采用这两个功能可以帮助提高患者的安全性,并进一步有助于质量保证。