Naturally occurring molecular species used for plasma diagnostics and signal correction in microwave-induced plasma optical emission spectrometry

机译:天然存在的分子种类,用于微波诱导等离子体光发射光谱法中的等离子体诊断和信号校正

获取原文
获取原文并翻译 | 示例
           

摘要

In the present study, we evaluate the N_2~+/OH emission intensity ratio as a diagnostic tool for identifying the best instrumental operating conditions in microwave-induced plasma optical emission spectrometry (MIPOES). This molecular species signal ratio is compared with the traditional Mg ll/Mg l ratio. Aluminum, Ba, Mn, Sr and Zn (analytes), and high concentrations of C, Na, Ca, HNO_3 and HCl (sample matrices) are used as models to investigate the effects of complex matrices on analyte recoveries. The N_2~+/OH signal ratio is more sensitive to changes in plasma conditions than the Mg ll/Mg I ratio. Some other advantages include real-time monitoring capabilities, and the possibility of independently tracking variations in both plasma and sample introduction. For less complex matrices, the N_2~+/OH signal ratio may be used for instrument optimization, which ensures plasma conditions are as similar as possible when analyzing standard solutions and samples. For analyses involving severe matrix effects, molecular species such as CN, N_2, N_2~+ and OH are used for signal correction. Significant improvements in accuracy are achieved by employing the analyte-to-molecular species signal ratio, or their product, for calibration. Both the use of the N_2~+/OH signal ratio as a diagnostic tool, and of molecular species for signal correction to minimize matrix effects are simple strategies that may significantly contribute to expanding the analytical capabilities of MIP OES and facilitating its application in routine analysis.
机译:在本研究中,我们评估N_2〜+ / OH的发射强度比,作为在微波诱导等离子体发射光谱(MIPOES)中识别最佳仪器操作条件的诊断工具。将该分子种类的信号比率与传统的Mg ll / Mg l比率进行比较。铝,钡,锰,锶和锌(分析物)以及高浓度的碳,钠,钙,HNO_3和盐酸(样品基质)被用作模型,以研究复杂基质对分析物回收率的影响。与Mg ll / Mg I的比率相比,N_2〜+ / OH信号的比率对血浆条件的变化更敏感。其他一些优势包括实时监控功能,以及独立跟踪血浆和样品引入变化的可能性。对于不太复杂的基质,可以将N_2〜+ / OH信号比用于仪器优化,以确保在分析标准溶液和样品时血浆条件尽可能相似。对于涉及严重基质效应的分析,将分子种类(例如CN,N_2,N_2〜+和OH)用于信号校正。通过使用分析物与分子种类的信号比率或其乘积进行校准,可以显着提高准确性。使用N_2〜+ / OH信号比作为诊断工具以及使用分子种类进行信号校正以最大程度地减少基质效应都是简单的策略,可能会极大地扩展MIP OES的分析能力并促进其在常规分析中的应用。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号