Multicentre use of a porous tantalum monoblock acetabular component

机译:多中心使用多孔钽整体髋臼组件

获取原文
获取原文并翻译 | 示例
       

摘要

The purpose of this study was to evaluate the minimum five-year prospective results from the multicentre use of a porous tantalum monoblock acetabular component for primary total hip arthroplasty (THA). A multicentre study was performed in 253 consecutive primary THAs in three separate surgical centres. All patients underwent identical postoperative protocols including radiological and clinical evaluation. The average preoperative total HHS score was 44.0 ± 13.8 and increased at one-year follow-up to 95.2 ± 4.8 (p <0.05), remaining constant through the five-year follow-up at 97.0 ± 6.2 (p < 0.05). There was no radiographic evidence of gross polyethylene wear, progressive radiolucencies, osteolytic lesions, acetabular fracture, or component subsidence. From these results, we can recommend the continued use of this material for acetabular components in primary THA and that further review of the current multicentre population is warranted to determine the long-term durability of the acetabular composite.
机译:这项研究的目的是评估多中心使用多孔钽整体式髋臼组件进行原发性全髋关节置换术(THA)的至少五年的预期结果。在三个独立的手术中心对253个连续的原发性THA进行了一项多中心研究。所有患者均接受相同的术后方案,包括放射学和临床评估。术前平均HHS总评分为44.0±13.8,在一年的随访中增加到95.2±4.8(p <0.05),在五年的随访中保持恒定在97.0±6.2(p <0.05)。没有影像学证据表明聚乙烯总磨损,进行性放射透进,溶骨性病变,髋臼骨折或下陷。从这些结果中,我们可以建议继续将该材料用于原发性THA中的髋臼组件,并且有必要进一步审查当前的多中心人群,以确定髋臼复合材料的长期耐用性。