首页> 外文期刊>International Journal of Thermophysics >The New LMK Primary Standard for Dew-Point Sensor Calibration: Evaluation of the High-Range Saturator Efficiency
【24h】

The New LMK Primary Standard for Dew-Point Sensor Calibration: Evaluation of the High-Range Saturator Efficiency

机译:露点传感器校准的新LMK主要标准:高量程饱和器效率的评估

获取原文
获取原文并翻译 | 示例
           

摘要

In the field of hygrometry, a primary dew-point standard can be realized according to several proven principles, such as single-pressure (1-P), two-pressure (2-P), or divided flow. Different realizations have been introduced by various national laboratories, each resulting in a stand-alone complex generation system. Recent trends in generator design favor the single-pressure principle without recirculation because it promises theoretically lower uncertainty and because it avoids problems regarding the leak tightness of the recirculation. Instead of recirculation, the efficiency of saturation, the key factor, is increased by preconditioning the inlet gas entering the saturator. For preconditioning, a presaturator or purifier is used to bring the dew point of the inlet stream close to the saturator temperature. The purpose of the paper is to identify the minimum requirements for the preconditioning system and the main saturator to assure efficient saturation for the LMK generator. Moreover, the aim is also to find out if the preconditioning system can be avoided despite the rather simple construction of the main saturator. If this proves to be the case, the generator design can be simplified while maintaining an accurate value of the generated dew point. Experiments were carried out within the scope of improving our existing primary generator in the above-ambient dew-point range up to +70°C. These results show the generated dew point is within the measurement uncertainty for any dew-point value of the inlet gas. Thus, the preconditioning subsystem can be avoided, which leads to a simplified generator design.
机译:在湿度测定领域,可以根据一些公认的原理来实现主要的露点标准,例如单压(1-P),双压(2-P)或分流。各个国家实验室已经引入了不同的实现方式,每个实现都产生了一个独立的复杂生成系统。发电机设计的最新趋势支持无再循环的单压原理,因为它从理论上保证了较低的不确定性,并且避免了与再循环的密封性有关的问题。通过预处理进入饱和器的进气,可以提高饱和效率(这是关键因素),而不是进行再循环。对于预处理,使用预饱和器或净化器使进口流的露点接近饱和器温度。本文的目的是确定预处理系统和主饱和器的最低要求,以确保LMK发生器的有效饱和。此外,目的还在于发现尽管主饱和器的结构很简单,还是可以避免使用预处理系统。如果事实证明如此,则可以简化发电机设计,同时保持所产生露点的准确值。在高于+ 70°C的上述露点范围内改进现有的初级发电机的范围内进行了实验。这些结果表明,对于进口气体的任何露点值,生成的露点都在测量不确定度之内。因此,可以避免预处理子系统,这导致简化的发电机设计。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号