Screen-printed electrodes incorporated in a flow system for the decentralized monitoring of lead, cadmium and copper in natural and wastewater samples

机译:丝网印刷电极集成在流量系统中,用于分散监测天然和废水样品中的铅,镉和铜

获取原文
获取原文并翻译 | 示例
           

摘要

Mercury-based screen-printed electrodes (SPE) combined with square-wave anodic stripping voltammetry (SWASV) techniques for the analysis of copper, cadmium, lead, and zinc in different water samples have been applied. The detection system has been implemented in a flow cell and different experimental conditions have been tested in view of its application for in-situ monitoring. In particular, an acetate buffer together with a low chloride concentration (0.025 M NaCl) provided best performance and reproducible results. Additionally, the flow system was validated for the first time in terms of limits of detection, linearity, repeatability and recovery. Limits of detection of 2.8 μg L~(-1), 4.1 μg L~(-1) and 7.5 μg L~(-1) for cadmium, lead and copper respectively and repeatabilities lower than 10% (as RSD) were found. Good recoveries have been obtained for the three cations and in particular for copper, even in the presence of zinc. Finally, the method has shown its efficiency for the rapid screening of lead, cadmium and copper contained in both natural waters and wastewater samples.
机译:基于汞的丝网印刷电极(SPE)与方波阳极溶出伏安法(SWASV)结合用于分析不同水样中的铜,镉,铅和锌。该检测系统已在流通池中实施,鉴于其在现场监测中的应用,已经测试了不同的实验条件。特别是,乙酸盐缓冲液与低氯化物浓度(0.025 M NaCl)一起提供了最佳性能和可重现的结果。此外,该流动系统首次在检测限,线性,可重复性和回收率方面得到验证。镉,铅和铜的检出限分别为2.8μgL〜(-1),4.1μgL〜(-1)和7.5μgL〜(-1),重复性低于10%(相对标准偏差)。即使存在锌,三种阳离子,尤其是铜也获得了良好的回收率。最后,该方法显示了其快速筛查天然水和废水样品中所含铅,镉和铜的效率。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号