A coaxial calorimeter and its use as a reference standard in an automated microwave power calibration system

机译:同轴量热仪及其在自动微波功率校准系统中用作参考标准

获取原文
获取原文并翻译 | 示例
           

摘要

The construction of a dc-substitution coaxial calorimeter having 7-mm-diameter input line is described and an evaluation of its uncertainty is presented. An automated procedure is then outlined for calibrating a power transfer standard against the calorimeter. The transfer standard consists of a directional coupler with a thermistor mount on its side arm. The power transfer standard can in turn be used for automated calibration of thermistor mounts and other power sensors. The overall estimated uncertainty (expressed as one standard deviation) of the measurement of the effective efficiency of a thermistor mount varies from 0.2 percent at 2 GHz to 0.4 percent at 18 GHz.
机译:描述了具有7mm直径输入线的dc替代同轴量热仪的结构,并对不确定度进行了评估。然后概述了自动程序,以针对量热计校准功率传输标准。传输标准包括一个定向耦合器,其侧臂上带有一个热敏电阻安装座。功率传输标准又可以用于热敏电阻安装座和其他功率传感器的自动校准。热敏电阻安装件的有效效率测量的总体估计不确定度(表示为一个标准偏差)在2 GHz时为0.2%,在18 GHz时为0.4%。