摘要

A novel oscillator circuit with automatic capacitance compensation (ACC) capability has been coupled with quartz-crystal-microbalance (QCM) sensors coated with quinoxaline- and pyrazine-bridged cavitands to detect aromatic and chlorinated compounds in water. With double-side immersed 10-MHz crystals coated with the quinoxaline cavitand, the detection of toluene in deionized water was possible down to concentrations of a few parts per million. The ACC oscillator advantageously provides the simultaneous measurement of the sensor resonant frequency, damping, and value of the compensated parallel capacitance. This enabled observing that the analyte sorption in the cavitand coating not only brings about a mass uptake but also an increase of losses and, apparently, a rise in the coating average permittivity.
机译:一种新型的具有自动电容补偿(ACC)功能的振荡器电路已与涂有喹喔啉和吡嗪桥空分子的石英晶体微天平(QCM)传感器结合使用,以检测水中的芳香族化合物和氯化化合物。使用涂有喹喔啉空泡孔的双面浸入式10 MHz晶体,可以检测低至百万分之几浓度的去离子水中的甲苯。 ACC振荡器有利地提供传感器谐振频率,阻尼和补偿的并联电容值的同时测量。这使得可以观察到,空洞涂层中的分析物吸附不仅引起质量吸收,而且损失增加,并且显然涂层平均介电常数增加。