Doppler angle and flow velocity estimations using the classic and transverse Doppler effects

机译:使用经典和横向多普勒效应估算多普勒角和流速

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Current clinical Doppler ultrasound systems could only measure the flow vector parallel to the ultrasound beam axis, and the knowledge of the Doppler angle (beam-to-flow angle) is needed to calculate the real flow velocity. Currently, the Doppler angle is determined visually by manually aligning a vessel axis marker along the blood vessel on the duplex scan image of the ultrasound. The application of this procedure is often limited by practical constraints; therefore, measurements are not reliable. In order to overcome this problem, the authors developed a simple Doppler angle and flow velocity estimation method using a combination of the classic and transverse Doppler effects. This method uses only a single focused annular array transducer to estimate the Doppler angle and the flow velocity. The authors have verified experimentally that this method is successful for measuring constant flow in a flow phantom between 45/spl deg/ and 80/spl deg/ Doppler angle. The standard deviation of the estimated Doppler angles is less than 4.5/spl deg/. This method could be implemented easily in medical Doppler ultrasound systems to automatically estimate the Doppler angle and the flow velocity.
机译:当前的临床多普勒超声系统只能测量平行于超声束轴的流动矢量,并且需要多普勒角(射束与流动角)的知识来计算实际流速。当前,通过在超声的双工扫描图像上沿着血管手动对准血管轴标记来视觉确定多普勒角度。该程序的应用通常受到实际限制。因此,测量不可靠。为了克服这个问题,作者开发了一种结合经典和横向多普勒效应的简单多普勒角度和流速估计方法。该方法仅使用单个聚焦环形阵列换能器来估计多普勒角度和流速。作者已通过实验验证,该方法可成功测量45 / spl deg /和80 / spl deg /多普勒角度之间的流动模型中的恒定流量。估计的多普勒角度的标准偏差小于4.5 / spl deg /。该方法可以在医疗多普勒超声系统中轻松实现,以自动估计多普勒角度和流速。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号