Doppler angle and flow velocity estimations using the classic andtransverse Doppler effects

机译:使用经典和横向多普勒效应估算多普勒角度和流速

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Current clinical Doppler ultrasound systems could only measure thenflow vector parallel to the ultrasound beam axis, and the knowledge ofnthe Doppler angle (beam-to-flow angle) is needed to calculate the realnflow velocity. Currently, the Doppler angle is determined visually bynmanually aligning a vessel axis marker along the blood vessel on thenduplex scan image of the ultrasound. The application of this procedurenis often limited by practical constraints; therefore, measurements arennot reliable. In order to overcome this problem, the authors developed ansimple Doppler angle and flow velocity estimation method using ancombination of the classic and transverse Doppler effects. This methodnuses only a single focused annular array transducer to estimate thenDoppler angle and the flow velocity. The authors have verifiednexperimentally that this method is successful for measuring constantnflow in a flow phantom between 45° and 80° Doppler angle. Thenstandard deviation of the estimated Doppler angles is less thann4.5°. This method could be implemented easily in medical Dopplernultrasound systems to automatically estimate the Doppler angle and thenflow velocity
机译:当前的临床多普勒超声系统只能测量平行于超声束轴的血流矢量,因此需要了解多普勒角(束流角)来计算实际流速。当前,通过手动地在超声的双扫描图像上沿着血管对准血管轴标记来在视觉上确定多普勒角。该程序的应用通常受到实际限制。因此,测量不可靠。为了克服这个问题,作者开发了一种结合经典和横向多普勒效应的简单多普勒角和流速估计方法。该方法仅对单个聚焦环形阵列换能器进行估计,然后估计多普勒角和流速。作者已通过实验验证了该方法成功地测量了在45°至80°多普勒角度之间的流动模型中的常数流。则估计的多普勒角度的标准偏差小于n4.5°。该方法可在医疗多普勒超声系统中轻松实现,以自动估计多普勒角度和流速

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号